operation manual

Specifications
Power supply 3.7V 420mAH built-in
rechargeable li-polymer
battery
Display Display with white back
light
Visible area = 46.8mm (L)
x29.3mm (W)
Measurement method Oscillometric method
Measurement range Rated cuff pressure:
0kPa-40kPa
(0mmHg-300mmHg)
Measurement pressure:
4kPa-34kPa
(40mmHg-230mmHg)
heart rate: 40-199 beats
per minute
Accuracy Pressure:
5°C-40°C with
in ±0.4kPa (±3mmHg)
heart rate:±5% of mea
surement result on dis
play
Normal operating
conditions
Temperature: 5°C to
40°C. Relative humidity:
85%RH. Atmospheric
pressure: 86kPa to
106kPa
Storage and trans
portation conditions
Temperature: -20°C to
60°C. Relative humidity:
10% to 93%. Atmospheric
pressure: 50kPa to
106kPa
Measurement perime
ter of the wrist
About 13.5cm-21.5cm
Net weight Approx. 100g
External dimensions Approx.
80mm×65mm×13.2mm
Accessories USB cable, user manual
Mode of operation Continuous operation
Degree of protection Type BF applied part
Protection against
ingress of water
IP22, This means: pro
tected against access to
hazardous parts with a
finger and against verti
cally falling water drops
when tilted up to 15
degrees.
Device classification Battery Powered Mode:
Internally Powered ME
Equipment. Class II ME
Equipment
Caution: No modification of this equipment is
allowed.
Electromagnetic emissions
and immunity
The device is approved according to EMC safety
standard EN 60601-1-2. It is designed to be used in
typical domestic environments.
EMC Guidance
- The Blood Pressure Monitor needs special pre
cautions regarding EMC and needs to be
installed and put into service according to the
EMC information provided in the accompanying
documents.
- Wireless communications equipment such as
wireless home network devices, mobile phones,
cordless telephones and their base stations,
walkie-talkies can affect this equipment and
should be kept at least a distance d = 3.3 m
away from the equipment.
Note: As indicated in IEC 60601-1-2:2007 for ME
equipment, a typical cell phone with a maximum
output power of 2 W yields d = 3.3 m at an immuni
ty level of 3V/m.
Guidance and manufacturer's declara
tion – electromagnetic emissions - for
all ME equipment and ME systems
Guidance and manufacturer’s declaration – elec
tromagnetic emissions
The device is intended for use in the electromag
netic environment specified below. The customer
or the user of the device should assure that it is
used in such an environment.
Emissions test Com
pliance
Electromagnetic
environment - guid
ance
RF emissions
CISPR 11
Group
2
The device must emit
electromagnetic
energy in order to
perform its intended
function. Nearby
electronic equipment
may be affected.
RF emissions
CISPR 11
Class B
Harmonic emis
sions IEC
61000-3-2
Not
applica
ble
Voltage fluctua
tions/flicker
emissions IEC
61000-3-3
Not
applica
ble
Guidance and manufacturer's declara
tion – electromagnetic immunity – for
all ME equipment and ME systems
Guidance and manufacturer’s declaration – elec
tromagnetic immunity
The device is intended for use in the electromag
netic environment specified below. The customer