General description
Proven tube-in-shell membrane technology is the basis of the SDS™-Series Drying System, which
selectively removes water vapor from a gas stream. The driving force is the vapor pressure
differential between the sample gas and a purge gas counter-flowed outside the tubing. This
difference drives the reaction, quickly drying the gas stream.
The SDS-Series drying system is designed as an add-on to an existing conditioning system. It can
also be used as the primary gas drying device, provided the sample is non-condensing at ambient
temperature.
Specifications and features
Sample flow rate:
o 4C-D201 5 lpm maximum
o 4C-D202 10 lpm maximum (single stream)
5 lpm maximum each (dual stream)
Sample inlet humidity: Non-condensing at room temperature
Sample outlet humidity: < -15°C dew point at maximum inlet flow and humidity conditions
Sample inlet pressure: 80 psig maximum
Sample inlet temperature: 90°C maximum
Dimensions: 24” x 6” x 6”
Enclosure: NEMA 12, IP 54, gray powder-coated finish
Purge air requirements:
o Purge gas pressure: 30-150 psig
o Purge gas humidity: less than -40°C dew point
o Oil-free instrument air or nitrogen recommended
Connections:
o Sample inlet: ¼” Kynar™ compression fitting
o Sample outlet: ¼” Kynar compression fitting
o Purge air inlet : ¼” female NPT
o Purge air outlet: ¼” female NPT; allows for connection to outside vent