Manual

PERMA PURE LLC
Model M5 User’s Manual
Doc. #580: Revision: 000
Page 6 of 22
Physical Description
Single channel system
2 x 10” heat exchangers connected in series only
1 Active (cooled to 5°C) heat exchanger
1 Passive heat exchanger
LCD temperature display and LED status indicators
Dimensions: 14.5” x 11” x 10.5” HWD (37 x 28 x 26.5 cm)
Weight: 53 lbs (24 kg)
Specifications
Sample gas flow range Up to 8 LPM (17 scfh)
Water content 20% H
2
O (60°C DP) @ 5 LPM
Maximum cooling rate 440 BTU/ht (464 kJ/hr)
Maximum inlet sample temp. 400°F (205°C) for SS or glass impingers
280°F (138°C) for Kynar™ impingers
Maximum inlet pressure 15 psig (1 bar)
Maximum pressure drop <+1 in H2O
Ambient temperature range 33-114°F (0.6-43°C)
Outlet sample gas dew point 41°F (5°C)
Voltage 115VAC (230VAC optional)
50/60 Hz
Power supply 1000W
Cooling down time Less than 10 minutes
Connections
Sample gas inlet: Stainless Steel, 1/4” tube compression fitting
Sample gas outlet: Stainless Steel, 1/4” tube compression fitting
Drain tubing connection: Kynar, 1/4” barbed tube fitting