P E R MA P U R E INSTRUCTION MANUAL THERMO-ELECTRIC COOLERS MODULAR SERIES Model M325 Perma Pure LLC 8 Executive Dr, PO Box 2105 Toms River, NJ 08754 www.permapure.com Tel: 732-244-0010 Tel: 800-337-3762 (toll free US) Fax: 732-244-8140 Email: info@permapure.
TABLE OF CONTENTS A: Specifications................................................................................................................ 3 B: Limited Warranty........................................................................................................... 4 C: Principle of Operation ................................................................................................... 5 D: Installation .........................................................................................
A: SPECIFICATIONS Physical Description Single (Series) / Dual (Parallel) Channel System 2 x 5” Heat Exchangers connected in series or parallel 1 Passive (cooled to ambient), 1 Active (cooled to 4°C) Heat Exchangers LCD temperature display Operating Specifications Sample Gas Flow Range 2-4 LPM 4.3-8.6 SCFH 168°F @ 20% 40% H LPM H2O 3LPM 2O, 4@ Inlet Dew Point at Rated Flow Maximum Cooling Rate 222 BTU/Hr Dimensions 11.20 x 7.25 x 11.20 in. HWD ° 76 C 234kJ/Hr 28.5 x 18.4 x 28.5 cm Weight 17 lbs 7.
B: LIMITED WARRANTY Perma Pure LLC WARRANTY and DISCLAIMERS Perma Pure (Seller) warrants that product supplied hereunder shall, at the time of delivery to Buyer, conform to the published specifications of Seller and be free from defects in material and workmanship under normal use and service.
C: PRINCIPLE OF OPERATION Thank you for purchasing a Perma Pure Baldwin™-Series Thermo-Electric Cooler. This Modular Series features a unique slim design leaving additional space to install or access other sample conditioning system components. A unique drop-down door on the Modular Series provides easy access to electronic boards and the power supply. All electronic boards (control, relay, and display) are mounted on the door for easy access.
316L stainless steel), PVDF (Kynar), or glass, is mounted within a thermally insulated heat transfer block bored to receive the heat exchanger without a mechanical lock. This assembly allows the easy removal of any heat exchanger simply by slipping it out of the cooling block by hand. The heat transfer block cools the heat exchanger through the heat pumping action of the peltier element.
D: INSTALLATION The Model M325 should be installed away from heat sources in a well ventilated area of an instrument rack or enclosure. The more stable the ambient temperature environment around the Model M325, the better the output dew point stability. Sample tubing connections to the Model M325 depend on the heat exchanger material of construction. A cooler with a stainless steel heat exchanger uses a stainless steel inlet fitting.
E: START-UP PROCEDURE Plug in the power cord to a properly grounded main circuit. The Ready Green LED will come on within 3 minutes, indicating the relay temperature (10°C) has been achieved. After approximately 3 minutes, the set point of +5°C. (41°F) is achieved. The SLIP Green LED should also be on if a water slip sensor is connected to the internal relay board. The sample gas flow may be started immediately after the READY Green LED comes on. The Model M325 is virtually maintenance free.
F: LED INDICATORS The Model M325 has two green and one red LED operating indicators. These indicators are arranged vertically on the front of the cooler. The bottom green LED indicates the READY operating temperature status, normally set for 10°C (50°F). After the set-point temperature is reached, the sample pump may be turned on by other devices. When the impinger temperature is below 10°C (50°F) the ready LED will be on. Above this temperature, the ready LED will be off.
G: I/O TERMINAL BLOCK DESCRIPTION The I/O terminal blocks are found on the bottom panel of the cooler: TB1 TB1 is the standard analog output (low voltage DC output) for all Modular Series Thermo-Electric Coolers. Model M325 is a single-channel cooler with one passive and one active 5” heat exchanger connected in series. The output is 0vdc to 2.5vdc for a temperature range of 0°C to 25°C. C Terminal 1 is earth ground.
H: TEST & ADJUSTMENT PROCEDURES NOTE: All test and adjustment procedures have been performed at the factory. Therefore, no adjustment should be necessary. A. Main Control board (1) WARNING: Before connecting power to the cooler, be aware of the HAZARDOUS LIVE VOLTAGE on the control board. Disconnect power from the cooler before opening the swing-down "L-door". Note: The auxiliary output board must be removed if the cooler is so equipped. Remove the thermocouples from TB1 and TB2.
B. Display Board (1) WARNING: Before connecting power to the cooler, be aware of the HAZARDOUS LIVE VOLTAGE on the control board. Disconnect power from the cooler before opening the swing-down "L-door". Remove display cover from the front of the swing-down "L-door". Connect a voltmeter between the ground test point, TP7 on the main control board and TP1 on the display board, the red lead to TP1 and the black lead to TP7.
For Current Output (3a) If the auxiliary analog output board is set up for current output, set the ammeter to 200ma range. Connect the ammeter to TB1 (the channel 1 current output terminal), the black lead to the negative (-) terminal and the red lead to the positive (+) terminal. (4a) Adjust POT4 (the zero pot) to obtain a 4mA reading on the ammeter. (5a) Set the thermocouple generator to 25°C. (6a) Adjust POT1 (the span pot) to obtain a 20mA reading on the ammeter.
I: DESCRIPTION OF OPTIONS Baldwin Modular Series Thermo-Electric Coolers have three available options: (1) temperature display; (2) alarm relay/water slip, and (3) auxiliary analog output. Modular Series Coolers may be equipped with any one or all of the options. All external I/O connections for these options are available through the terminal blocks on the bottom of the cooler. 1. Display Option (standard on model M325) The display option is a secondary board that is mounted on the main control board.
J: “NEW JERSEY” THERMOCOUPLE OPTION Some air quality management districts (e.g., those in New Jersey and Southern California) require temperature measurement of the gas stream at the outlet of the last heat exchanger on the cooler. Perma Pure offers a 1/32-inch diameter hypodermic-style type K thermocouple that can be inserted into a special heat exchanger (i.e., it has a small port for insertion of the thermocouple) so the actual sample dew point temperature can be measured.
K: TROUBLESHOOTING Symptom No LED(s) and no fan. Check AC power input. No LED(s) and fan on. AC input fuse on control board. DC output fuse on control board. VC on control board. Action Ensure that AC power is connected. Replace fuse as necessary. Replace fuse as necessary. Replace control board. LED(s) on and no fan. AC input fuse on power supply. +12vdc TB4 on control board. Replace fuse as necessary. Replace power supply. Impinger remains at ambient temperature. Thermocouple failure LED is on.
For further service assistance, contact: Perma Pure LLC P.O. Box 2105 8 Executive Drive (08755) Toms River, NJ 08754 Tel: 800-337-3762 (toll free U.S.) Tel: 732-244-0010 Fax: 732-244-8140 Email: info@permapure.
L: SPARE PARTS Model M325 Part No.
APPENDIX A: MODEL M325 19
Modular Series Thermo-Electric Coolers Control Board
Modular Series Thermo-Electric Coolers Display Board
Modular Series Thermo-Electric Coolers Auxiliary Analog Output Board
A B C D TB2 1 2 3 100 R4 C4 CHN1_READY TC_1_OK WATER_SLIP2 WATER_SLIP1 CHN2_READY TC_2_OK C6 .1uf 1 2 3 4 5 6 7 15uf 3 2 1 C7 15uf 1 3 6 2 4 11 8 12 10 VCC C18 .1uf C23 .1uf C24 .01uf 1 2 3 C13 .01uf VCC C14 .1uf YEL RED TB1 Channel 1 P1 C2 15uf + + C5 .1uf LM556C CON TRG THRES RST U4B LM556C CON TRG THRES RST U4A 100 R3 C3 C8 .1uf C1 .1uf 1 2 3 4 5 6 7 9 13 NOTE 5 5 1 .
A B C D 2 1 1 2 3 4 5 6 7 8 VCC VCC CHN1_READY TC_1_OK WATER_SLIP2 WATER_SLIP1 CHN2_READY TC_2_OK VCC VCC 1 1 1 1 100K RN1D 100K RN1C VCC 100K RN1B 100K RN1A 5 4 3 2 1 2 NOTE 1: For single sample streams only, if separate water slip and temperature alarms are required, jumper pins 1 & 2 on JP1, JP2 and JP3. For dual sample streams, jumper pins 3 & 4 on JP1, JP2 and JP3. Water Slip 1 TB5 2 1 P1 Water Slip 2 TB2 C8 .1uf C3 .1uf C7 .1uf C6 .001uf C4 .1uf C5 .
APPENDIX B: SAMPLE CONDITIONING SYSTEM 20
AIR DIMENSIONS INCORPORATED 1371 West Newport Center Dr., Suite 101, Deerfield Beach, FL 33442 - Phone 954-428-7333 or 800-423-6464 Fax 954-360-0987 http://www.airdimensions.com e-mail address -Info@AirDimensions.com MINI DIA-VAC® MAINTENANCE AND DISASSEMBLY INSTRUCTIONS A. General Operations Characteristics 1. Normal motor coil temperatures may be 160 - 180 degrees F. Winding insulation is Class B. Please note the two fans are different, so before removing the fans, note which side they belong on. 2.
*Diaphragms require close precision tolerance, therefore only ADI diaphragms should be used as replacements. C. Disassembly of Head Section and Service Diaphragm 1. Remove head section by unscrewing the four large bolts. A flat-bladed screw driver may be needed to gently pry the head free of the service diaphragm. **If you have Teflon coating on the heads use caution not to scratch the surface. 2.
E. Related Torque Values 1. Head bolts - 110 inch pounds. 2. Valve body screws and Diaphragm plate screws - 70 inch pounds Dia-Vac® is a Registered Trademark of Air Dimensions Inc.
%( / +P*I 25+) 5KPING .
1. Single Pump Head Loading Note: Use only MASTERFLEX Precision Tubing with MASTERFLEX Pumps to insure optimum performance. Use of other tubing may void applicable warranties. Contents: One pump head, one 15 in (38 cm) length of silicone tubing, one mounting hardware package, manual and tubing loading key. Supplied tubing loading key required for assembly. a) Separate the end bells (the pump head halves). Hold the end bell containing the rotor as shown with the tubing retainer grooves facing down.
Set contains four #8-32 screws, four washers, and four wingnuts. Number of heads To be mounted 1 2 3 4 Cold- rolled steel Order number MN-07013-02 MN-07013-03 MN-07013-03 MN-07013-07 Stainless steel Order number MN-07013-04 MN-07013-05 MN-07013-08 MN-07013-09 2. Multi-Channel Mounting Flat bladed screwdriver required for mounting. Tubing loading key required for mounting. Note: Other special mounting hardware for multi-channel pumping. See “ 3. Replacement Parts and Accessories”.
CHART OF VOLUME PERCENT WATER CONCENTRATIONS AT SATURATION FOR VARIOUS TEMPERATURES AT STANDARD PRESSURE (ATMOSPHERIC PRESSURE) DEGREES C DEGREES F VOLUME % DEGREES C DEGREES F VOLUME % +100 + 212 100.00 +2 + 36 0.696 + 90 + 194 69.20 +1 + 34 0.649 + 80 + 176 46.70 0 + 32 0.602 + 75 + 167 38.70 -1 + 30 0.555 + 70 + 158 30.70 -2 + 28 0.510 + 65 + 149 25.20 -3 + 27 0.469 + 60 + 140 19.70 -4 + 25 0.431 + 55 + 131 15.50 -5 + 23 0.396 + 50 + 122 12.20 -6 + 21 0.363 + 45 + 113 9.45 -7 + 19 0.333 + 40 + 104 7.
MOISTURE CONVERSION TABLE DEWPOINT F C -110 -108 -106 -104 -102 -100 -98 -96 -94 -92 -90 -88 -86 -84 -82 -80 -78 -76 -74 -72 -70 -68 -66 -64 -62 -60 -58 -56 -54 -52 -50 -48 -46 -44 -42 -40 -38 -36 -34 -32 -30 -28 -26 -24 -22 -20 -18 -16 -14 -12 -10 -8 -6 -4 -2 0 +2 +4 +6 +8 +10 +12 +14 +16 +18 +20 +22 +24 +26 +28 -166 -162 -159 -155 -152 -148 -144 -141 -137 -134 -130 -126 -123 -119 -116 -112 -108 -105 -101 -98 -94 -90 -87 -83 -80 -76 -72 -69 -65 -62 -58 -54 -51 -47 -44 -40 -36 -33 -29 -26 -22 -18 -15 -1