Owner`s manual
Specifications 50
7
7 Specifications
Power
Requirements
Voltage 120VAC (NTSC MODEL) 230VAC (PAL MODEL)
Frequency 50 - 60 Hz
Input current 2.8 A 1.4 A
Voltage fluctuation ±10%
Operating
Environment
Ambient temperature 10 - 40°C
Relative humidity 30 - 85% (with no condensation)
Air pressure 700 - 1060 hPa
Storage
Environment
Ambient temperature –20 - 60°C
Relative humidity 0 - 85% (with no condensation)
Air pressure 700 - 1060 hPa
Illumination
Lamp 150-W twin halogen lamp
Average life span 50 h
Color temperature 3,400 K
Brightness control Selection: Automatic/Manual
Scope Compatibility
PENTAX color video
endoscopes
This processor can be used with
- PENTAX color video endoscopes (K-series) except 30K, 40K,
70CIMK/CIFK/CILK (Confocal Endomicroscopes) and 70UK/
UTK/URK (Ultrasound Endoscopes) series
NTSC MODEL PAL MODEL
PENTAX fiberscopes
This processor can be used with
- PENTAX fiberscopes with use of PENTAX video adapter
module PVK-1070 and PVK-1070Z
Air Feed System
Air pump system DC Diaphragm
Pressure setting at flow rate at 0 45 - 72 KPa (maximum pressure)
Standard air feed volume at inlet
of water bottle
Low: 2.0 - 4.0 L/min
High: 4.1 - 7.0 L/min
Water Feed System
Water bottle assembly
pressurized by pump
Bottle capacity = 250 ml
Brightness
Control System
Automatic Selection: Average / Peak
Manual ±5 step adjustment
Color System Color correction Red/Blue ±5 step adjustment
Freeze Function
Live video is displayed in a sub-image while freeze mode is
activated.
Cooling Forced air cooling
Video Signals
Digital output SERIAL 1 set 1 set
Analog output
RGB 1 set
Y/C 2 sets
Composite 1 set
Y 1 set
Control Signals
Remote 2 sets
Keyboard 1 set
Classification as
Electro Medical
Equipment
Type of protection against
electric shock
Class I equipment, 3 prong plug
Degree of protection against
electric shock
Type BF (Body Floating), using insulated endoscope.
Use on a heart is prohibited.
Degree of explosion proofing
Do not use in potentially flammable surroundings.*
* The processor is not suitable for use in a mixture of air and
flammable anesthetic gas or a mixture of oxygen/nitrous oxide and
flammable anesthetic gas.
Electro Magnetic
Compatibility
Designed in accordance with
EN60601-1-2 (2001) + A1 (2006) for EU
IEC60601-1-2 (2001) + A1 (2004) for other countries
Compliance Designed in accordance with UL60601-1 IEC60601-1
Size Excluding protrusions Width 380 mm × Height 155 mm × Depth 430 mm
Weight Main Body 11.0 kg