J Error Messages
Message Meaning Solution
Reference
Page
Please replace the lamp 1
The lamp is not lit. Replace lamp 1 or lamp 2. P.25
Please replace the lamp 2
Please switch the lamp The lamp has gone out during use. Switch to the spare lamp. P.44
Scope is not Connected The scope is not attached. Attach the scope. P.18
Connected Scope is not
Compatible with this Processor
Please Contact with your
Distributor
The scope is not compatible. Use a K-series scope.
P.18
P.50
Scope Communication Error Communication with the scope failed. Reattach the scope. P.18
W/B Failed Please Retry The white balance is incorrect. Set the white balance again. P.27
Lamp Error
Please close Lamp Door
The lamp housing cover is open. Close the lamp housing cover. P.44
Lamp Switching Error Lamp is not positioned correctly.
Open the lamp housing cover and
check the harnesses are attached
correctly.
P.44 - 45