5 Maintenance and Storage
Follow the procedure below after using the processor.
1
Turn off the peripheral devices. For details on operating peripheral devices, refer to the
instruction manuals for the peripheral devices.
2
Press the power switch to turn off the processor.
3
Remove the scope and water bottle from the
processor. (Refer to Figure 5.1)
Remove the scope while keeping it straight and making sure not
to use excessive force on the connector part.
4
Disconnect the power cord.
5
Wipe the surfaces of the processor with a gauze that has been dampened with 70-90% ethyl or
isopropyl alcohol.
The procedures below should be followed to decontaminate surfaces of the EPK-p video processor that are suspected
of being contaminated.
1 Wipe surfaces with a gauze lightly dampened with an enzymatic detergent (neutral pH, low foaming, mild/compatible
with PENTAX products, etc.)
2 Wipe the same areas with a gauze lightly dampened with clean water (to remove residual detergent)
3 Wipe the processor with a dry gauze or lint-free cloth to remove remaining water/fluids
4 Wipe the surfaces of the processor with a gauze dampened with 70-90% ethyl or isopropyl alcohol
6
For details on after use care of the scope and water bottle, refer to the owner’s manuals for the
scope and water bottle.
CAUTION
Observe the following points on storage.
Be sure to turn off the processor and disconnect the power cord prior to storage.
Do not store the processor in a very hot and humid location, or where it is likely to be exposed to water or direct sunlight.
Store the processor in a location where dust will not enter inside. Wipe off any dust on the processor. Excessive amounts
of dust accumulating inside the unit may cause the processor to malfunction, emit smoke, or catch fire.
Take care not to drop the processor or subject it to an excessive mechanical shock. The processor may become unsafe to use
or damaged. If the processor is subject to an excessive mechanical shock, stop using it and contact our service personnel.
After Use Care and Storage
z Turning off the processor prior to turning off any connected peripheral devices may cause the peripheral devices to
malfunction.
z Make sure no water gets on the processor when you are performing the after care procedure. In particular, getting
the connector parts and ventilation grids wet may cause a malfunction.
2