Owner`s manual
Operating Procedure 41
4
Output the image to an external device.
1
Press the [Remote out] key.
• The image displayed on the monitor is output to a printer or other external output device.
Output the image to a PC.
1
Press the [Capture] key.
•
Output the image data displayed on the monitor to a PC with Image Capture Software (OS-I1) for the endoscope installed.
Set the menu color and enhancement.
1
Press the [Processor option] key.
• The current setting value appears.
2
Press the [I] or [J] key repeatedly to change the [Menu Color] setting to [Deep] or [Light].
• The current setting value appears.
Deep: Dark color
Light: Light color
3
Press the [I] or [J] key repeatedly to change the [Scope Enhancement] setting to [Disable] or
[Enable].
Disable: Use processor enhancement.
Enable: Use scope enhancement.
CAUTION
4
Press the [Enter] key or [Esc] key.
• Pressing the [Enter] key returns to the normal screen where any changes to the settings are saved and reflected.
• Pressing the [Esc] key returns to the normal screen where any changes to the settings are discarded.
Set the beep tone.
1
Press the [Change beep tone] key.
• The beep tone setting switches between ON, OFF, and Alphabet OFF.
+ [Remote out] Key
+ [Capture] Key
+ + [Processor option] Key
If set to [Disable], be sure to set it back to [Enable] after use. Enhancement may not work properly when you connect
the scope you are using to another processor.
+ + [Change beep tone] Key