Owner`s manual

Operating Procedure 25
4
J Freeze Screen
While frozen, the scope video is displayed in a sub-image on the monitor.
The date and time are hidden when the sub-image is displayed. You can change the display position of the sub-image in the
[Display] settings.
I "Settings in the [Display] Tab" (P.35)
CAUTION
1
Press c.
The processor lamp and lamp indicator will illuminate. After the lamp lights, any scope connected to the processor will
emit light out of its distal end.
CAUTION
Front Panel Operations
z Do not use a sharp object such as the tip of a pen to press the switches on the front panel.
z Do not use the end of the scopes' light guide to operate the keys or switches.
Lamp Switch
z About the Lamp Life:
The lamp life for the EPK-p processor is 50 hours. Check the service lamp life indicator on the front panel before you
use the processor. If the service light is orange for either service light 1 or 2 and "Please replace lamp (1 or 2)"
appears on the monitor, replace the affected lamp.
I "Replacing the Lamp" (P.44)
z If the lamp does not light:
If the lamp does not light when you press the c, LAMP button, check the service light lamp indicators on the front
panel and replace any bad lamps by referring to "Replacing the Lamp" (P.43).
z To turn off the lamp:
After use, be sure to press c to turn the lamp off.
Sub-
Image
Main Screen