18 Preparation
2
Connect a scope to the processor. Use a scope with a water bottle and suction unit connected to the scope. You can connect
a PENTAX K-series video endoscope (P.50) to the processor. For details on the scope, refer to the owner’s manual for the
scope.
1
Align the scope locking lever of the processor to the
OPEN position. (Refer to Figure 2.1)
2
Insert the electrical connector and light guide of the
scope into the corresponding scope electrical
connector and light guide attachment of the
processor and move the scope locking lever in the
direction of
2 until it clicks into the LOCK position.
3
Attach water bottle assembly, 2/3 filled with sterile water, to the appropriate location on the left
side of the video processor.
NOTE
• Align the A/W-DRAIN lever to the A/W (air/water supply)
position. (Refer to Figure 2.2)
• Insert the air/water connector into the holder of the cap until the
scope is connected. (Refer to Figure 2.3)
Connecting an Endoscope
z Do not over tighten the cap.
z Replace the sterile water every day.
z Be sure to use sterile water.
1
2
Scope
Locking
Lever
Electrical
Connector
Light Guide
Figure 2.1
A/W
DRAIN
Figure 2.2
Figure 2.3