Owner`s manual

Preparation 15
2
Install the processor and peripheral devices while referring to “System Configuration” (P.14). Install them on a stable and level
surface.
CAUTION
J Connecting an Isolation Transformer
Connect the processor and peripheral devices to an isolation transformer. To reduce the possibility of electric shock connect
the video processor and peripheral devices to the PENTAX SAT-1300 Isolation Transformer or another Medical Grade Isolation
Transformer of the appropriate power rating.
1
Connect the power cord of the isolation transformer to a three-prong power outlet that meets the
power rating indicated on the rating plate.
2
Use the AC power cord to connect the power input socket on the rear panel of the processor to
the isolation transformer.
Make sure the processor is turned off beforehand.
3
Connect the peripheral devices to the isolation transformer.
Make sure the peripheral devices are turned off prior to installation.
For details on peripheral devices, refer to the instruction manuals for the peripheral devices.
CAUTION
Installation
z Do not install the processor in any of the following locations.
- Where the processor is likely to be exposed to water.
- Where flammable or explosive gas is present.
- In very hot and humid locations.
- Where the processor will be exposed to direct sunlight.
z Do not install the processor vertically. Doing so may cause a malfunction or water to leak.
z Install the processor in a location where the ventilation grids will not be blocked.
z Install the processor in a location where dust will not enter inside. Wipe off any dust on the processor. Excessive
amounts of dust accumulating inside the unit may cause the processor to malfunction, emit smoke, or catch fire.
z Do not hold the scope locking lever when you move the processor.
z Do not connect any electrical devices other than the processor and peripheral devices of the processor to the isolation
transformer.
z Make sure the voltage, current, and power consumption of the processor and peripheral devices do not exceed the
maximum ratings indicated on the isolation transformer.
z Be sure to securely connect the power cord of the isolation transformer to a three-prong power outlet.
Isolation Transformer
Processor
Peripheral Device
AC Power Cord