Manual

13
The thumb button of the plunger is made of anodized aluminum. The syringe
and plunger should never be soaked in bleach
(Clorox™or Clidox™) as this
may cause permanent corrosion and/or staining.
The FMJ-250 High Pressure Syringe can be autoclaved. It should not be heated
above
standard autoclave temperatures of 105 - 135 degrees Centigrade.
The FMJ-250 High Pressure Syringe may also be cleaned with organic
solvents, very hot water, or sonicated.
Always dry the syringe, put the protective the black end cap on the device before storing
and in the original protective plastic box provided on shipment.
Replacement plungers for FMJ-250 High Pressure Syringe
After long use or from over-pressuring the system, the white Teflon® tip of the syringe plunger
may become worn. One symptom of this is that the syringe may leak from its proximal end (around
the plunger rod). If this happens, a replacement plunger may be purchased from Penn-Century,
Inc. without having to replace the entire syringe. Contact us for details.
ALWAYS test both devices with water if they have been stored and not used for a long period
of time.
For a full summary of the company’s repair and warranty policies, please review the Penn-Century Terms
and Conditions.
As manufacturers, Penn-Century, Inc. can offer:
Product knowledge
Operation of our devices
Familiarity with published studies related to use of the same device and, if possible, similar drugs
Familiarity with customer feedback about them
These instructions and photos are the property of the Penn-Century, Inc. They must be not be
reproduced without the written consent of Penn-Century, Inc.
IMPORTANT NOTICE: Requirements for references to Penn-Century devices in publications,
abstracts and presentations
As a condition of sale, the Buyer agrees that upon publication or presentation of any research conducted
using a Product, the Penn-Century name as the manufacturer ("Penn-Century. Inc. Wyndmoor, PA") and
the name of the Products must be correctly spelled and cited, and include Penn-Century’s registered
trademark and model number as follows and as applicable: “MicroSprayer® Aerosolizer Model ___,
“FMJ-250 High Pressure Syringe”, “Dry Powder Insufflator™ - Model ___” or “Small Animal Laryngoscope
Model LS-2”. A breach of this section shall result in immediate harm to Penn-Century, and in addition
to any and all rights available to Penn-Century, Penn-Century shall be entitled to an injunction or other
equitable relief as a remedy for such breach.
For any questions, contact us at Penn-Century, Inc. info@penncentury.com or 215-753-6540.
Rev. 14Nov2012