User Manual

4.Front Panel Operation
17
Series
CH1 and CH2
in series
CH3
The connection method of the series connection of CH1 and CH2 is as shown in the
figure below.
CH2CH1
V
2
R
L
V
1
V
L
V
L
= V
1
+ V
2
POWER SUPPLY
Channel track
In independent output mode, set the output parameters of CH1 and CH2, and then
enter the channel track mode, if the parameters of any one channel are changed, the
other channel will change proportionally.
For example, in independent output mode, set the CH1 voltage to 2V, current to 1A;
set the CH2 voltage to 4V, current to 2A. After entering channel track mode, if CH1
voltage is set to 6V, CH2 voltage will be adjusted to 12V proportionally. If CH1 current
is set to 2A, CH2 current will be adjusted to 4A proportionally.
Note: If the setting value is out of the rated range, it will be set to the maximum.
To set the output mode:
(1) Press the Utility key. The Output mode menu is selected.
(2) Press the / key to select the output mode. Press the key to check and
enter the selected mode.