User Manual

4.Front Panel Operation
16
is only for CH1 and CH2, without affecting CH3. There are four output modes:
Independent Output
The parameter of each channel can be set independently.
Parallel track
When CH1 and CH2 are connected in parallel, you can select this mode to simplify
the parameter inputting. You just need to set the parameters of the combined
channel. The voltage rating is same as the single channel; the current rating is the
sum of CH1 and CH2 current rating.
Press the orange ON/OFF key to turn on/off the combined channel.
Parallel
CH1 and CH2
in parallel
CH3
The connection method of the parallel connection of CH1 and CH2 is as shown in the
figure below.
CH2CH1
I
2
R
L
I
1
I
L
I
L
= I
1
+ I
2
POWER SUPPLY
Series track
When CH1 and CH2 are connected in series, you can select this mode to simplify the
parameter inputting. You just need to set the parameters of the combined channel.
The voltage rating is the sum of CH1 and CH2 voltage rating; the current rating is
same as the single channel.
Press the orange ON/OFF key to turn on/off the combined channel.