User Manual

4.Front Panel Operation
11
Press the orange Curr/CC key, the first digit of the CH1 set current is flashing. Press
the key, the first digit of the CH1 O.C.P is flashing, indicating the value is editable.
There are two methods to change the value.
Modify: Turn the knob to change the value. Press the
/
key to move the
cu rso r. Press the knob or the key to confirm.
Input: Use the numeric keys to input, the input box of Channel 1 limit current will
pop up. Enter a desired value. Press the key to confirm.
Set the O.C.P of CH2
Press the blue Curr/CC key, then press the key to enter edit mode. You can set the
value in the same way as CH1 above.
Set the O.C.P of CH3
Press the Curr CH3 key, then press the key to enter edit mode. You can set the
value in the same way as CH1 above.
4.4 Programmable Output
The programmable output function can preset up to 100 groups of timing parameters.
When you turn on the programmable output, the instrument will output the
pre-specified voltage, current in pre-specified time.
4.4.1 Data View
Press the Program key. The Data view menu is selected.
(1) The Memory sub menu is selected. Press the
/
key to select Internal or
External.
(2) Press the key to select Import submenu. Press the key to import data.
(3) Press the key to select Export submenu. Press the key to export data.
Note: When the memory is set as External, the programmable data file will be
export to U disk, the directory is P6181\Program.
(4) Press the key to select Clear Data submenu. Press the key to clear data.
4.4.2 Output Set
Press the Program key, turn the knob to select [Output Set].
(1) The Cycle Mode sub menu is selected. Press the
/
key to select Oder or
Loop.
(2) Press the key to select Start Point submenu. Use the numeric keys to input (1 to
100), press the key to confirm.