User Manual
4.2. Calibration 
1.)  Pour  a  small  quantity  of  ph  6.86,  ph  4.01  and  ph  9.18  solution  into 
clean beakers. 
2.)  For a particularly accurate calibration, it is advised to use two beakers 
for each buffer solution; the first is to be used for rinsing the electrode, 
the second is  to be used for the calibration.  In this way, the risks of 
contaminating the buffer solution are reduced to a minimum. 
PH7 
3.)  Switch on the instrument. 
4.)  Immerse the pH electrode into a standard buffer solution with  
pH = 6.86. 
5.)  Stir slowly the electrode in the solution until you get a stable reading. 
6.)  Use  a  small  screwdriver  (supplied)  to  adjust  the  calibration  trimmer 
(PH7) at the front of the instrument until the display shows “6.86”.  
7.)  Rinse the electrode with distilled water. 
PH4 
8.)  Immerse the pH electrode into a standard buffer solution with  
pH = 4.01. 
9.)  Stir slowly the electrode in the solution until you get a stable reading. 
10.) Use  a  small  screwdriver  (supplied)  to  adjust  the  calibration  trimmer 
(PH4) at the front of the instrument until the display shows “4.01”.  
11.) Rinse the electrode with distilled water. 
PH9 
12.) Immerse the pH electrode into a standard buffer solution with  
pH = 9.18. 
13.) Stir  slowly  the  electrode  in  the  solution  and  wait  for  the  reading 
stabilized until it displays 9.18. 
14.) Rinse the electrode with distilled water. 
15.) The calibration for the instrument is now complete. 
-16- 










