User Manual

4. PH Calibrating Procedure
4.1. Calibrating Consideration
This pH meter was calibrated by a mV signal, which was simulated from the
ideal mV query of the pH electrode (on the basis of an environment of 25°C).
Has to be considered:
a) An ideal electrode produces 0 mV at a value of pH = 7.0, however,
most electrodes differ slightly.
b) The temperature of the measurement environment is not always about
25°C, but mostly between 15°C and 35°C.
Under certain circumstances it is necessary to carry out the following
calibration procedure to adjust the electrode of the instrument to the
maximum accuracy.
Important:
The instrument’s pH range must be re-calibrated whenever:
The electrode has been used (laid) for long time from the last calibration
The electrode has been used in particularly taxing conditions.
The utmost accuracy is required.
After replacing the new electrode
-15-