User Manual
Light:
Measurement range: 20...20000 Lux
(20000 Luxrange = reading x 10)
Overrate Display: Highest digit of „1“ is displayed
Accuracy: ± 5% rdg. + 10 digit (Calibrated to
standard incandescent lamp at colour
temperature 2856 k)
Repeatability: ± 2 %
Temperature-
Characteristics: ± 0,1 % / °C
Photo-detector: One silicon photo diode with filter
Humidity:
Measurement range: 25 %..95 % RH
Resolution 0,1 % RH
Accuracy: ± 5% (at 25°,35...95% RH)
Input Protection: 60 V DC or 24 V AC
rms
Temperature:
Measurement range: -20..+200°C , +200..+750°C
-4…+200°F, +200…+1400°F
Resolution: 0,1, 1°C
Accuracy: ± 3% rdg. ± 2°C
(-20...+200°C/-4…+200°F)
± 3,5% ± 2°C
(-20...+750°C/+200…+1400°F)
Input Protection: 60 V DC or 24 V AC
rms
-17-