Datasheet
8.9 General Characteristics 
Interface 
Isolated USB Host interface 
Copy saved file to PC from a U disk, then analyze it with upper computer 
software. 
Isolated LAN interface 
For remote control of the Analyzer and measurement data transmission. 
Screen 
Color TFT LCD 
Size 
5.6 inch 
Resolution 
320×240 
Brightness 
Adjustable 
Memory 
Flash memory 
128M 
TF card 
Standard 8Gbyte 
Standards 
Measurement method 
IEC61000-4-30 S degree 
Measurement performance 
IEC61000-4-30 S degree 
Power quality monitoring 
EN50160 
Flicker 
IEC61000-4-15  
Harmonic 
IEC61000-4-7 
Environment 
Working temperature 
0°~ 40° 
Storage temperature 
-20°~ 60° 
Humidity 
< 90% relative humidity 
Safety 
Complied with 
IEC 61010-1 
Safety Degree: 1000V CAT III / 600V CAT IV 
Pollution Degree: 2 
Maximum input voltage 
600V CAT IV / 1000V CAT III 
Maximum voltage at current 
input 
42Vpk 
Mechanical 
Dimension 
262× 173 × 66 mm 
Weight 
1.6 kg 
Power 
Adapter input 
90 ~ 264V 
Adapter output 
12V / 2 A 
Battery 
Rechargeable Ni-MH battery 7.2V 3.8Ah 
Battery operating time 
>7 hours 
Battery charge time 
5 hours 
Specifications of optional current clamps 
Model 
Range 
Turns ratio 
Accuracy 
Size (mm) 
P 4145-5A 
5A 
10mV/A 
±0.2% 
8mm 
P 4145-50A 
50A 
10mV/A 
±0.2% 
8mm 
P 4145-100A 
100A 
1mV/A 
±0.2% 
13mm 
P 4145-1000A 
1000A 
1mV/A 
±1.0% 
52mm 
P 4145-3000A 
3000A 
65mV/1000A 
±1.0 (±2% position error) 
160mm 
-74- 