Datasheet
8.6 Measuring Modes and Parameters 
Measurement mode 
Measured parameters 
Oscilloscope 
Vrms, Arms, Vcursor, Acursor, Hz 
Voltage/current/frequency 
Vrms, Vpk, Vcf, Arms, Apk, Acf, Hz 
Dips&Swells 
Vrms1/2, Arms1/2, capture up to 1000 events, include date, time, 
duration, magnitude and phase mark, and threshold is 
programmable settable. 
Harmonic 
1-50, harmonic voltage, THD voltage, harmonic current, THD 
current, interharmonic voltage, interharmonic current 
Power and energy 
W, VA, VAR, power factor, shift power factor, Arms, Vrms, kWh, 
kVah, kVARh 
Flicker 
Pst (1 minute), Pst, Plt, PF5 
Unbalance 
Vneg, Vzero, Aneg, Azero, Vfund, Afund, Hz, V phase angle, A 
phase angle 
Transient 
Vrms, Vcursor 
Inrush current 
Inrush current, Inrush duration, Arms1/2, Vrms1/2 
System monitoring 
Vrms, Arms, Harmonic voltage, Total harmonic distortion voltage, Plt, 
Vrms1/2, Arms1/2, Vneg, Hz, Swells, Dips, Interruption, Rapid 
Voltage Change. All parameters are measured simultaneously 
according to EN50160 standard. 
Specify unreliable readings according to IEC61000-4-30 standard. 
Logger 
User-defined to select more parameter and record of setting time 
interval 
8.7 Measurement Range, Resolution, Accuracy 
Voltage / Current / 
Frequency 
Measurement 
range 
Resolution 
Accuracy 
Vrms (AC+DC) 
1~1000 Vrms 
0.1 Vrms 
±0.5% of nominal voltage 
Vpk 
1~1400 Vpk 
0.1 Vpk 
±0.5% of nominal voltage 
V(CF) 
1.0~>2.8 
0.01 
±5% 
Arms (AC) 
10mV/A 
1mV/A 
50mV (65mV)/1000A 
0 ~ 100 A 
1 ~ 1000 A 
15 ~ 5000 A 
0.1 A 
0.1 A 
1 A 
± 0.5% ± 0.2 A 
± 0.5% ± 0.2 A 
± 1.0% ± 2 A 
A (CF) 
1~10 
0.01 
±5% 
Frequency 50Hz nominal 
Frequency 60Hz nominal 
Frequency 400Hz nominal 
42.5~57.5 Hz 
51~69 Hz 
340 ~ 460 Hz 
0.01 Hz 
0.01 Hz 
0.01 Hz 
±0.01 Hz 
±0.01 Hz 
±0.01 Hz 
Dips&Swells 
Measurement 
range 
Resolution 
Accuracy 
Vrms1/2 
0~200% nominal 
voltage 
0.1Vrms 
±1% 
Arms1/2 
1~3000A 
1A 
±1%±2A 
Threshold value 
Threshold is settable according to nominal voltage percentage 
Detectable events type: Dips, Swells, Interruption, Voltage Rapid 
Change. 
Duration 
hour-minute-second-
microsecond 
0.5 cycle 
1 period 
Flicker 
Measurement 
range 
Resolution 
Accuracy 
Flicker severity (Pst, Plt) 
and the instantaneous 
flicker sens. level 
0.00~20.00 
0.01 
Within ±5% of table value 
according to IEC61000-4-15  
-72- 