Datasheet
 Events table 
The events table shows the events that occurred during the measurement with time of start, phase and 
duration. Events recording situation: 
V rms events: an event is recorded each time that a 10 minute aggregated RMS value violates its limits.  
Harmonics events: an event is recorded each time a 10 minute aggregated harmonic or THD violates its 
limit. 
Dips/Interruption/Rapid Voltage Change/Swells events: an  event  is recorded each time one of  the 
items violates its limit. 
Unbalance and Frequency events: an event is recorded each time that a 10 minute aggregated RMS 
value violates its limit. 
Available function keys: 
F3 
Access Trend Screen 
F4 
Switch between selected and all events. 
F5 
Return to previous menu. 
 Trend 
Press【F3】 in Events Table to access Trend Screen. Press left/right arrow keys to access horizontal 
time base zoom. 
Available function keys: 
F5 
Return to previous menu. 
 Bar Graph Screen 
The main system monitor display shows the worst harmonic for each of the three phases. Function key 
【F2】 brings up a screen with Bar Graphs showing the percentage of time each phase spent within 
limits  for  25  harmonics  and  Total  Harmonic  Distortion  (THD).  Each  Bar  Graph  has  a  wide  base 
(representing an adjustable limit of e.g. 95%) and a narrow top (representing the limit of 100%). A Bar 
Graph changes from green to red if the limits for that harmonic are violated. 
With the left/right arrow keys you can position the Cursor on a particular Bar Graph and measuring data 
belonging to that bar is shown in the screen header. 
Available function keys:  
F1 
Selection of Bar Graphs belonging to phase A (L1), B (L2) or C (L3). 
F4 
Access to events table. 
F5 
Return to main menu. 