7.6. Power and Energy 
Power  and  Energy  displays  a  table  with  all  important  power  parameters.  The  related  Trend  screen 
shows the changes over time of all measuring values in the table. 
 Table 
The table displays power data for each phase and in total: real or active power (kW), apparent power 
(kVA,  the  product  of  rms  voltage  and  current),  reactive  power  (kVAR,  the  reactive  component  of 
apparent power caused by phase shift between AC current  and  voltage in inductors and capacitors), 
power  factor  (TPF,  the  ratio  of  real  power  to  apparent  power  for  the  total  rms  including  harmonics), 
displacement power factor  (DPF, the ratio of real power to apparent power for fundamental), and the 
rms values of voltage and current. 
A popup table with energy usage by phase and in total can be activated by pressing 【F3 】 –Energy 
function key. The table shows real energy (kWh), apparent energy (kVAh) and reactive energy (kVARh). 
The energy measurement starts when Power&Energy is started. The readout can be reset with function 
key F5. 
Available function keys: 
F3 
Access the Energy popup screen. 
F4  
Access the Trend screen. 
F5 
Switch between RUN and HOLD. 
Energy popup screen 
F3 
Close Energy 
F4 
Access the Trend 
F5 
Reset, the time clock starts all over again 
 Trend 
The figures in the table are instantaneous values that update constantly. While changes in these values 
over time are recorded in the Trend. The traces build up from the right side. The readings in the header 
correspond to the most recent measurements plotted on the right. 
Available function keys: 
F1 
Switch parameter being displayed. 
F4 
Return to the Table screen. 
F5 
Switch between RUN and HOLD. 