5.2. First use 
At delivery, the built in chargeable battery may be empty and it is recommend to charge it before use. A 
full charge for the first time takes at least 6 hours, it means the battery is fully charged when the charging 
indicator change from red to off. No damage will occur if the adapter is connected with the Analyzer for 
long periods. The Analyzer automatically cuts off charging when the battery is fully charged. Before use, 
check  that  the  adapter  voltage  and  frequency  range  match  the  local  line  power  range.  To  prevent 
decrease of batter capacity, charge it at least twice a year. 
Before measurements, set the Analyzer according to the line voltage, frequency and wiring configuration 
of power system you want to measure, read section ”Set the Analyzer” for more information.  
5.3. Inputs 
Picture: Connection to a 3-Phase system 
The  Analyzer  has  4  BNC-inputs for  current clamps  and 5  banana-inputs for  voltages.  For  a  3-phase 
system, make the connections as shown in the figure. 
First put the current clamps around the conductors of phase A (L1), B (L2), C (L3) and N (neutral). The 
clamps are marked with an arrow indicating the correct signal polarity.  
Next make the voltage connections: start with Ground and then succession N, A (L1), B (L2), C (L3). For 
correct measuring results, always connect the Ground input. 
For  single  phase  measurements,  use  voltage  input  A  (L1)  or  current  input  A  (L1)  or  voltage  inputs 
Ground. 
-43- 