User Manual

15.3. I/O Settings
The analyzer supports communication trough the LAN, USB and RS232 interface.
1. I/O interface
Enalbe the LAN, USB and RS232, or disable the all.
2. LAN
Set the information for LAN.
3. USB
Set the information for USB.
4. RS232
Set the information for RS232
15.4. Ref Source
Select the internal or external reference source, the default is internal.
16. Source
The menu key is used to enable or disable the tracking generator, also to set the amplitude of
tracking signal output. The source function is only valid for the analyzer with tracking generator option.
16.1. Source
To set the tracking generator output to be On or Off. Default is Off.
16.2. Amplitude
To set the amplitude of tracking generator output.
Normalize
Connect the GEN OUTPUT 50Ω output of TG with the RF INPUT 50Ω input of the analyzer. Then
press Normalize key to eliminate the measurement error caused by its frequency response.
17. Measure
The analyzer provides advanced measurement function including Channel Power, Adjacent Channel
Power and Occupied Bandwidth.
17.1. Channel Power
Measures the total power of signal within the specified channel bandwidth, which includes the
Channel Power and Power Density.
Channel power: the power within the integral bandwidth
Power Density: the power normalized to 1Hz within the integral bandwidth.
Select Channel Power measurement under Meas function and press Meas Setup to set
corresponding parameters.
17.2. Integ BW
Specify the range of integration used in calculating the power in the channel
-41-