User Manual

9.14. Marker Table
Open or close the marker table. When the table opened, all opened markers are listed on table
and shown on the lower window of screen, including Marker number, Marker trace number,
readout type, X-axis reading and amplitude. User can watch several measurements of points and
up to 8 markers at the same time.
Note: The opened table can be saved in the external memory and recalled when needed. User
can press Save key to store the data, detail method is descriped in “Save”.
Figure 2-9
9.15. All Off
Turn off all the markers and related functions.
9.16. Freq Count
Frequency counter reading
Figure 2-10
1. On/Off
Turn on or off the frequency counter.
Key points:
If no active marker selected, turing on the frequency counter will open a Normal marker
automatically.
When frequency counter opened, the frequency readouts will be more accurate.
In Zero span mode, the frequency counter is enabled to measure the frequency of center
frequency nearby.
-34-