6.4. Points
Set the desired points for each sweep. That is the points of current trace.
Key points:
When sweep time is limited by smapling rate of ADC (Analog to Digital Converter), modifying
the sweep numbers will affect sweep time. That is, the more pionts, the longer sweep time wil
be.
Modifying the numbers also influences other system parameters, so sytme will restart sweep
and measure.
You can modify this parameter by numeric keys, knob and arrow keys. For more details, please
refer to “Parameter Input”.
Table 2-16 Sweep pionts
6.5. SWT Count
Set the number of time for a single sweep. The system will execute specified numbers of sweep and
the number in icon shown at the left status bar of screen changes along with the process of sweep.
Table 2-17 Sweep Count
-26-
Parameter
Description
Default
601
Range
101 ~ 3001
Unit
No
Knob Step
1
Arrow Keys Step
100
Hint
Along with the increasing of the sweep points (more than 601), the resolution
of marker point will be increased too, but the sweep speed will be decreased.
Except of Sweep Points, other parameters also influence the speed of sweep,
Like Span, RBW, VBW, Detector Type and Center Frequency.