4.9. Scale/Div
Only in Log scale mode, set the logarithmic units and scale value per vertical grid division on the
display.
Key points:
Trough changing the scale, the displayed amplitude range is adjusted.
Amplitude range can be displayed:
Min:Reference level – 10 × scale ;
Max:Reference level
You can modify this parameter by numeric keys, knob and arrow keys. For more details, please
refer to “Parameter Input”.
Table 2-8 Scale
4.10. Scale Type
Set the scale type of Y-axis to Lin or Log, the default is Log.
Key points:
In Log scale type, Y-axis denotes the logarithmic coordinate, top grid shows reference level, and
the grid size is equal to the scale value. The unit of Y-axis will be switched to be dBm as default
when the scale type is changed from Lin to Log.
In Lin scale type, Y-axis denotes linear coordinate, top grid shows reference level and bottom
grid is 0V, the grid size is10% of the reference level and scale setting is invalid. The unit of Y-
axis will be switched to be Volts as default when the scale type is changed from Log to Lin.
Scale type will not affect the unit of Y-axis.
-18-
Parameter
Description
Default
10dB
Range
0.1 dB ~ 20 dB
Unit
dB
Knob Step
Scale≥1, step=1 dB
Scale<1, step=0.1 dB
Arrow Keys Step
In 1-2-5 sequence