Datasheet

Basic Settings
3.1. FREQ
Set the frequency parameter of analyzer. The analyzer sweeps within a specified range, and the
sweep will start once you change this parameter.
The frequency range of current channel can be expressed by either of two groups of
paramenters: Start Frequency/Stop Frequency f
start
/f
stop
, or center Frequency/Span
f
center
/f
span
. If any one of them is adjusted, other three parameters will be changed
accordingly to ensure their coupling relationship.
f
center
=f
start
-f
stop
/2
f
span
= f
stop
- f
start
3.2. Center Frequency
Enable the center frequency function, to allow a frequency displays in center be selected. When
pressed, the frequency mode is switched to Center Frequency/Span for input. In this mode,
parameters you specified are always shown at the lower right and left side of the display grid.
Key points:
The start and stop frequencies vary with the center frequency when span is constant.
Changing the center frequency horizontally shifts the current sweep channel and the
adjustment is limited by the specified range in datasheet.
In the Zero Span mode, start frequency, stop frequency and center frequency are always
equal.
User can modify this parameter by numeric keys, knob and arrow keys.
Table 2-1 Center Frequency
Parameter
Description
Default
0.75GHz (P4130-1)
1.1GHz (P4135-1)
1.5GHz (P4140-1)
Range
0Hz ~ 1.5GHz (P4130-1)
0Hz ~ 2.2GHz (P4135-1)
0Hz ~ 3.0GHz (P4140-1)
Unit
GHz, MHz, kHz, Hz
Knob Step
Span>0, step=span/200
Span=0, step=RBW/100, Min=1 Hz
Arrow Keys Step
CF step