User Manual
6
3. :FUNCtion Commands
The control commands of waveform function are as follows:
1. :FUNCtion
Syntax
:FUNCtion?
:FUNCtion
{SINE|SQUare|RAMP|PULSe|NOISe|ARB|DC|AM|FM|PM|FSK|PWM|S
WEep|BURSt}
Function
Query/set the waveform function for current channel.
Explanations
This command work on the current selected channel by default. If you
want to set the other channel, you need switch channel first (refer to the
command of channel control, such as CHANnel CH2).
Example
:FUNCtion RAMP
2. :FUNCtion:SINE:LOAD
Syntax
:FUNCtion:{SINE|SQUare|RAMP|PULSe|NOISe|ARB|DC}:LOAD?
:FUNCtion:{SINE|SQUare|RAMP|PULSe|NOISe|ARB|DC}:LOAD
{ON|OFF|<ohm>}
Function
Query/set the load of current channel.
Explanations
Ω is the unit of <ohm>, the default is 50Ω. "OFF" sets the output terminal
as “High Z”; "ON" sets it as the setting value.
{SINE|SQUare|RAMP|PULSe|NOISe|ARB|DC} have the public parameter
"LOAD". Only take SINE for instance here, other waveforms will not be
repeated.