5.2.7 Set DC offset
Press Offset/Low levsoftkey then set the desired offset value using the knob or numeric
keypad. Here, enter by knob is strongly recommend because of more convenience feature.
The DC offset setting will limit by amplitude and level, which should be agreed with below
formula:
Limit Low Vpp/2OffsetLimi High Vpp/2
If the specified offset is not valid, the waveform generator will automatically adjust it to the
maximum DC voltage allowed with the amplitude specified.
If set amplitude to 0V, high level limit to +10Vdc and low level limit to -10Vdc, then the
offset can be set within ±10V range. The waveform generator become one set of DC power
supply and can output DC voltage signal.
5.2.8 Set phase
Press Phase/Alignsoftkey to selectPhase then set the desired phase using the numeric
keypad or knob.
Output phase means the phase difference between output signal and synchronous signal,
and output signal advance to synchronous signal.
Press Phase/Align softkey to selectAlign to make the synchronous signal of CHA and
CHB are with same phase. So its easy to calculate the phase difference of two channels on
basis of the phase setting for CHA and CHB.