5.2.6 Set amplitude
You can set amplitude by amplitude or level. If amplitude set is selected, high level and
low level of signal will change at the same time when changing amplitude but DC offset
remain unchangeable. If level set is selected, no matter high level or low level changing,
low or high level remain unchangeable but DC offset will change. The relationships
between Vpp, High, Low and offset are shown below:
Vpp=High-Low High=Offset+Vpp/2 Low=Offset-Vpp/2
In continuous menu, press softkey Ampl/High lev to set amplitude or high level. Press
softkey Offset/Low lev to set low level.
Amplitude Limitation: Press Limit/Moresoftkey, then selectLimit High orLimit Low
to specify the limit value of high level or low level. Even if wrong operation that exceed the
limit value, the generator wont be damage and perform within the limit value.
But if specify the high level value to be +10Vdc and low level to be -10Vdc, the limit function
wont work anymore. Besides, the amplitude setting will be limit by DC offset, the amplitude
value (V
pp
) should be confined as below,
V
pp
2×(Limi HighOffset) V
pp
2×(Offset Limit Low)
Not only this, the maximum amplitude will also be limited by frequency (see chapter 5). If
V
pp
exceed the limitation of above formula or frequency, waveform generator will modify the
setting value to make it confined within the allowable maximum amplitude.
Influenced by channel bandwidth, the output amplitude will decrease if frequency is higher.
So flatness compensation will be needed. To improve the sweep speed, its not available
for frequency sweep, and the amplitude will decrease if span higher frequency.
For the arbitrary waveform generator, if V
pp
dont reach to full range, the display value is not
agreed with the output value.