Default Valuelocation stores the factory default calibration value, to prevent it from damage, which
is not allowed for store operation.
If calibration disabled, the generator is in safe state and can’t allow store operation.
5.22.8 Calibration Value Recall
Press Cal Recallsoftkey then select ‘Cal Recall’ to recall the calibration value from non-volatile
memory. Press Cal Recallsoftkey then two storage locations displays in operation menu, Default
Value and User Value. The location of Default Valuestores the factory default calibration
value.
If user stores a wrong value by wrong operation, the waveform generator will be out of work. In this
case, you must recall default calibration value from Default Valuelocation then store toUser
Value memory. The waveform generator will recall automatically the value fromUser Value
memory and applied to it.
5.22.9 Disable Calibration
There are two ways to disable calibration as below, which will make ‘Closed’ displayed and the
waveform generator operate in security conditions.
1. Store the calibration value, the calibration state will turn off automatically after finishing storage.
2. If you don’t want to store, press Passwordsoftkey then select ‘Calibrate’, input wrong password
(except of 6900) to disable calibration.
5.22.10 Exit Calibration
If you select the other mode during the calibration process, the instrument will remain the state of last
calibration. If the state is not you want, press Exitsoftkey, the waveform generator will resume the
state of power on.
5.23 System Reset
Press Utilitykey then to Resetsoftkey to recall the default setting value.