User Manual

128
amplitude nominal value. Make sure that perform the amplitude calibration from 5# to 7#.
5.22.6 Flatness Calibration
The amplitude of output signal will decrease as the frequency increasing if frequency is greater than
1MHz. Therefore, you need to calibrate for different frequency point. Amplitude flatness uses the
relative comparison method and makes the amplitude with 100kHz frequency as comparison
standard. The nominal amplitude is 14dBm and 0dBm.
(3) PressNextsoftkey until the calibration number becomes to 20#, then perform the calibration for
the first section amplitude. Calibration condition is Frequency=100kHz and Amplitude=14dBm.
Use spectrum analyzer to test the actual output amplitude as standard value of first section.
PressNextsoftkey again, the calibration number will plus one automatically. The calibration
condition is Frequency=01MHz but amplitude nominal value remain unchanged. Adjust the
calibration value to make the output amplitude equal to standard value of first section amplitude.
Make sure that perform the calibration of first section amplitude from 20# to 60#.
(4) PressNextsoftkey until calibration number turns to 70# to perform the flatness calibration for
second section. The condition is Frequency=100kHz and Amplitude=0dBm. Use the spectrum
analyzer to test the actual output amplitude and make this value as standard value.
PressNext softkey again, the calibration number plus one automatically. The condition is
Frequency=01MHz and amplitude remain unchanged. Adjust the calibration value to make the
actual output amplitude equal to standard value of second section. Make sure that perform the
calibration for second amplitude flatness from 70# to 110#.
5.22.7 Calibration Value Storage
Make sure to store the calibration value after finishing calibration in case that the value is lost once
power off. Press Cal Storesoftkey then select ‘Cal Store’ to store present calibration value in
non-volatile memory and does not change when power has been off. Store the new calibration value
will cover the previous value in memory location. To protect the previous calibration value, an enquiry
Will cover original data, Store?’ will pop up before storage. Only press Yes softkey to enable the
storage. Otherwise, press Cancelsoftkey.
Press Cal Storesoftkey then two storage locations displays in operation menu,Default Value
and User Value. Press User Valuesoftkey to store the present calibration value to appointed
storage location. ‘Store’ will be displayed after finishing storage.