User Manual

4.13.1. Enable calibration:
* After booting, the calibration is in off state, and the generator could not be calibrated without inputting
calibration password, this is a way to protect calibrated parameters which may be changed carelessly.
* To enable calibration, select sine wave and then pressShift】【Cal-key, the calibration password
displayed as 0, input calibration password 1900, press N-key to enable calibration.
4.13.2. Parameters calibration:
* Press Menu-key to display calibration value on the left, and calibration sequence number on the right
when setting calibration conditions automatically.
* Adjust calibration value to calibrate present selected calibration option and make the output expected.
* Continue to press Menu-key and the calibration sequence number will increase step by step, users
could calibrate all those options respectively, which is shown in the following list.
* During calibration process, pressCal-key at any time then press Menu-key to return the
calibration sequence number to 00.
Parameter calibration table
Sequence
No.
Default calibration
value
Output nominal
value
Adjust the calibration value till the output
is within the error range
00
2047
0 VDC
Zero calibration:
output DC voltage as: -20 ~ 20 mVDC
01
870
10 VDC
Offset calibration:
output DC voltage as: 9.88 ~ 10.12 VDC
02
873
7 Vrms
Amplitude calibration:
output AC voltage as: 6.928 ~ 7.072 Vrms
03
300
0,71 Vrms
Amplitude calibration:
output AC voltage as: 0,701 ~ 0,719 Vrms
04
500
1 MHz
Frequency calibration:
output frequency as: 1 MHz ± 20 Hz
05 ~ **
100
5 Vpp
Flatness calibration:
output amplitude as. 4,5 Vpp ~ 5,5 Vpp
** P 4055: Sequence-no.: 05~07; P 4060: Sequence-no.: 05~24
4.13.3. Disable calibration:
* After finishing the calibration, pressCalkey and there display 1900, press any numeric keys
thenNkey to store the calibration parameters, disable calibration and exit the process.
* During the calibration process, if wrong calibration occurred, press Freqkey at any time to disable
calibration and exit without storing calibration parameters.
* After rebooting, the generator automatically recalls and uses the calibration parameters stored during last
calibration.
4.14. Default setting
4.14.1. Continuity function:
continuity function is default after booting.
Waveform: sinewave
Frequency: 1 kHz
Amplitude: 1 Vpp
Attenuation: auto
Offset: 0Vdc
Duty cycle: 50%
-43-