Refer to next page and press 〖Insert Wave〗key to enter into the waveform selection
window. Select desired waveform and press〖Eneter〗key.
For your selected waveform, press 〖Waveform〗key to show a setting frame, which
allows the user to make parameter setting for the inserted waveform. Up/Down key for
select and knob for value setting, no option setting means default value.
(1) Amplitude: set the Vpp for inserted waveform
(2) Offset: set the DC offset voltage for inserted waveform
(3) Phase: set the start phase for inserted waveform, refer to the phase of which.
(4) Insert position: set the position (X-axis value) where the editing waveform you
want to be inserted.
(5) Total points: set the total points for inserted waveform. Generator abstract
points from inserted waveform at interval, which is also the waveform length of
inserted waveform.
When parameter setting finished, press 〖Return〗key to back editing window and you will
see the selected waveform locate at your desired position.
The example shows how to insert a sine wave with 200 points.
(1) press 〖Insert Wave〗key to enter into the waveform select interface. Then press
〖Normal Wave〗key and select Sine.
(2) Press 〖Enter〗to confirm the operation and then 〖Waveform〗key to display
parameter setting frame. Set amplitude to be 1.0Vpp, offset 0Vdc, phase 0°, insert
position 0, total points 200, as below picture:
(3) Keep press〖Return〗until back to edit waveform window and you will see the
inserted Sine as you set.