Datasheet
-69-
DC Current
Range
Resolution
Accuracy
Protection against overcharge
600.0uA
0.1uA
±(0.9%reading
+
5digits)
Quick fuse 800mA/1000V
6000uA
1uA
60.00mA
0.01mA
600.0mA
0.1mA
±(0.9%reading
+
8digits)
10.00A
0.01A
±(1.5%reading
+
8digits)
Quick fuse 10A/1000V
AC TRMS Current
Range
Resolution
Accuracy(*)(50Hz÷1kHz)
Protection against overcharge
600.0uA
0.1uA
±(1.2%reading
+
5digits)
Quick fuse 800mA/1000V
6000uA
1uA
60.00mA
0.01mA
600.0mA
0.1mA
10.00A
0.01A
±(1.5%reading
+
5digits)
Quick fuse 10A/1000V
(*) Accuracy specified from 5% to 100% of the measuring range, sine wave.
Accuracy PEAK function: ±10%rdg , AC+DC TRMS Current: accuracy (50Hz÷1kHz):
±(3.0%reading + 20dgt)
Flexible coil Current
Range
Resolution
(50Hz÷60Hz)
(61Hz÷1kHz)
Protection against overcharge
30.00A
0.01A
±
(
0
.
8
%
r
e
a
d
i
n
g
+
5
d
i
g
i
ts)
±
(
2
.
4
%
r
e
a
d
i
n
g
+
5
d
gt
)
1000VDC/ACrms
300.0A
0.1A
3000A
1A
Diode test
Function
Test current
Max voltage with open circuit
<1.5mA
3.3VDC
Resistance and Continuity test
Range
Resolution
Accuracy
Buzzer
Protection against overcharge
600.0Ω
0.1Ω
±(0.5%reading + 10dgt)
<50Ω
1000VDC/ACrms
6.000kΩ
0.001kΩ
±(0.5%reading + 5digits)
60.00kΩ
0.01kΩ
600.0kΩ
0.1kΩ
6.000MΩ
0.001MΩ
60.00MΩ
0.01MΩ
±(2.5%reading + 10dgt)