User Manual
3. Functions and Ranges
3.1. DC voltage measurement
Range
Resolution
Accuracy
Overload
protection
600 mV
0,1 mV
0,5% rdg + 5 dgt.
1000 V AC/DC
6 V
1 mV
0,8% rdg. + 5 dgt.
60 V
10 mV
600 V
100 mV
1000 V
1 V
1,0% rdg. + 2 dgt.
Input impedance: > 10 M
3.2. AC voltage measurement (True RMS)
Range
Resolution
Accuracy
Overload
protection
600 mV
0,1 mV
1,0% rdg. + 5 dgt.
1000 V AC/DC
6 V
1 mV
60 V
10 mV
600 V
100 mV
750 V
1 V
1,5% rdg. + 5 dgt.
Input impedance: > 10 M
Frequency: 40 Hz – 400 Hz
3.3. DC current measurement
Range
Resolution
Accuracy
Overload
protection
600 µA
0,1 µA
1,5% rdg. + 3 dgt.
0,63 A / 1000 V
6000 µA
1 µA
60 mA
0,01 mA
600 mA
0,1 mA
6 A
1 mA
1,5% rdg. + 5 dgt.
10 A / 1000 V
10 A
10 mA
-35-