User Manual
When  you  short  the  test  leads  in  the  400  Ω  range,  your    meter 
display  a  small  value  (no  more  than  0.3  Ω).  This  value  is  due  to 
your meter´s and test leads internal resistance. Make a note of this 
value and subtract it from small resistance measurements for better 
accuracy. 
5.8. Continuity check 
Warning: 
To  avoid  electric  shock,  never  measure  continuity  on  circuits  or 
wires that have voltage on them. 
1.  Set the function switch to the "OHM" position. 
2.  Insert  the  black test  lead  banana  jack  into  the  negative  COM 
jack and the red test lead banana-plug into the positive Ω jack. 
3.  Press the "MODE" push button until the "  )))" symbol appears 
in the display. 
4.  Touch  the  test  probe  tips  to  the  circuit  or  wire  you  wish  to 
  check. 
5.  If  the  resistance  is  less  than  150  ohms  (P  3355) or  35  ohms 
(P  3360),  the  audible  signal  will  sound.  The  display  will  also 
show the actual resistance. 
5.9. Diode test 
Warning: 
To avoid electric shock, do not test any diode that has voltage on it. 
1.  Set the function switch to " 
 " position. 
2.  Press the "MODE" button until the "   " symbol appears in 
the display. 