User Manual

3.4. AC current measurement
Range
Resolution
Accuracy
Overload
protection
400 µA
100 nA
1,5% + 5 dgt.
0,5 A/1000 V fuse
6 x 32 mm
4000 µA
1 µA
40 mA
10 µA
2,0% + 5 dgt.
400 mA
100 µA
10 A
10 mA
2,5% + 5 dgt.
10 A/1000 V fuse
10 x 38 mm
400µA ~ 400mA range:
Displays effective value of sine wave (mean value response).
10A-Range:
For continuous measurement 10 seconds and interval-not less
than 15 minutes.
Frequency: 40Hz 400Hz
3.5. Resistance Test
Range
Resolution
Accuracy
Overload
protection
400
0,1
1,2% + 2 dgt.
1000 V
p
4 k
1
1,0% + 2 dgt.
40 k
10
400 k
100
4 M
1 k
1,2% + 2 dgt.
40 M
10 k
1,5% + 2 dgt.
400 - 40M - range:
Open circuit voltage: approx. 3V
-38-