Note:
1. When Auto-LCR mode is active, the secondary parameter
will show the equivalent resistance in parallel mode (Rp) to
replace the D factor if the C measured value is less than
5pF.
2. In order to avoid damaging the instrument, the capacitance
requires discharge before measuring.
8.3. Resistance measurement
1. Turn on the power.
2. Insert the measured resistance into the input terminal
directly.
3. Connect the alligator clips to the ends of the measured
resistance.
4. The default test mode is Auto LCR mode, the resistance
value will show on primary display and the phase angle θ
will show on secondary display. In Auto LCR mode, the
D/Q/ESR key, SEL/PAL key, SORTING key and REL%
key are not available.
5. Press FUNC. key three times to select Auto-R (ACR)
mode. The primary LCD display will show the resistance
value. The secondary parameter is omitted and the
D/Q/ESR key is not available.
6. Press FREQ key to select frequency value:
100/120Hz/1/10/100 kHz.
7. Press SER/PAL key to select series or parallel mode.
8. Press FUNC. key four times to select DCR mode. The
primary LCD display will show the resistance value. The
secondary parameter is omitted and the D/Q/ESR key,
SEL/PAL key and FREQ key are not available.