User Manual
Press REL% key again to show the current readings on
primary display and the “REL” symbol will be blinking. The
percentage range is from -99.9% ~ 99.9%. When the present
value is larger than double of initial value, the “OL%” indication
will be shown on the secondary display.
During relative measurement, analog bar is always indicating
the present measurement value but not the relative value.
7.7. Open/Short calibration
1. Press CAL key larger than 2 seconds to start the
open/short calibration procedure.
2. In open calibration mode, the secondary display will show
“Open”. There are two ways for open state input:
3. When using square terminals, the square terminals and
L
CUR
/L
POT
/H
POT
/H
CUR
terminals hang in the air
4. When using L
CUR
/L
POT
/H
POT
/H
CUR
terminals, insert the black
and red testing lines with alligator clip into the "L
CUR
", "L
POT
"
terminal and "H
CUR
" , "H
POT
" terminal respectively
5. Press CAL key and the 30-second countdown will be
shown on LCD panels. If the open calibration is finished,
the PASS or FAIL symbol will shown on the primary
display. Press CAL key again to save the calibration data
and enter into the short calibration mode.
-54-