7. Operating instruction
7.1. Power on the meter
* Press On/Off key to turn on the power
* The default mode is AUTOLCR smart mode and the
default test frequency is 1 kHz
* When ON/OFF- key is pressed during power-on mode, the
instrument will enter power-off mode
* The LCD will show the “OFF” state before powering off
7.2. Parameters setting
Press FUNC. key to select the following parameters
sequentially: AUTO LCR, L-Q, C-D, R, DCR.
Parameter
Meaning
AUTO
LCR
Auto LCR smart mode
L-Q
Inductance measurement the parameter on
sub-display is quality factor Q.
C-D
Capacitance measurement, the parameter on
sub-display is dissipation factor D.
R
Resistance measurement
DCR
DC resistance measurement mode
L/C/R measurement readings can be positive or negative. In C-
D measurement, if the main parameter is "-", the actual
component being tested is inductive; In L - Q measurement, if
the main parameter is "-", the actual component being tested is
capacitive; Theoretically, R is positive, in some cases, R is "-" ,
which may be calibration error, please re-calibrate the
instrument.