User Manual
4. Functions and Ranges
4.1. DC voltage measurement
Range
Resolution
Accuracy
Overload protection
600 mV
0,1 mV
0,5% rdg. + 4 dgt
600 V DC/AC
p
6 V
1 mV
60 V
10 mV
600 V
100 mV
Input impedance: > 10 M
4.2. AC voltage measurement
Range
Resolution
Accuracy
Overload protection
600 mV
0,1 mV
0,8% rdg. + 10 dgt.
600 V DC/AC
p
6 V
1 mV
60 V
10 mV
600 V
100 mV
Input impedance: > 10 M
Displays effective value of sine wave (mean value response)
Frequency: 40 Hz – 400 Hz
4.3. DC current measurement
Range
Resolution
Accuracy
Overload protection
600 µA
0,1 µA
1,0% rdg. + 10 dgt.
0,6 A / 600 V
6000 µA
1 µA
60 mA
0,01 mA
600 mA
0,1 mA
6 A
1 mA
1,2% rdg. + 10 dgt.
10 A / 600 V
10 A
10 mA
10 A-Range:
Duration measurement: <5A
Measurements at >5A for maximal 2 minutes on load and 15 minutes off
load.
-33-