6.Use the Multimeter (Optional) 
104 
(1)  Select Hold in the bottom menu as ON. HOLD will be shown on the display.   
(2)  Select OFF to exit this mode.   
Making Relative Measurements   
When making relative measurements, reading is the difference between a stored reference 
value and the input signal.   
(1)  Select Configure in the bottom menu, select Relative in the right menu to enter the 
relative mode.   
The measurement value at this time  is stored as the reference value, and displayed 
behind  △.   
In this mode, current reading = input value - reference value. 
(2)  Press it again to exit the mode.   
Note: This function is not available when measuring resistance, diodes, and testing for 
continuity.   
Information Display 
Show/hide the information window on the top right of the display. 
(1)  Select Configure in the bottom menu, select Show Info in the right menu as ON. The 
information window
will be shown on the display.   
(2)  Select OFF to hide.   
Auto or Manual Range 
Auto range is set as default. To switch auto or manual range, do the following steps: 
(1)  Select Configure in the bottom menu.   
(2)  Select Auto Range in the right menu, AUTO will be shown on the display.   
(3)  Select  Switch Range  in the right menu,  MANUAL  will  be shown on the display. 
Press this softkey to switch range. 
Note: When testing diode, testing continuity, and measuring capacitance, manual range is 
disabled.   
Multimeter Recorder   
You can use the multimeter data recorder to record the measurements when measuring 
current/voltage by multimeter (optional).   
Push the Utility button, select Function in the bottom menu, select DAQ in the left menu. 
The description of DAQ Menu is shown as the follows: 