Phase difference=(Delay between channels at the rising edge÷Period)×360°.
Other measurements
+PulseCount : The number of positive pulses that rise above the mid reference
crossing in the waveform.
-PulseCount
: The number of negative pulses that fall below the mid reference
crossing in the waveform.
RiseEdgeCnt
: The number of positive transitions from the low reference value to the
high reference value in the waveform.
FallEdgeCnt
: The number of negative transitions from the high reference value to the
low reference value in the waveform.
Area : The area of the whole waveform within the screen and the unit is
voltage-second. The area measured above the zero reference (namely the vertical offset) is
positive; the area measured below the zero reference is negative. The area measured is the
algebraic sum of the area of the whole waveform within the screen.
Cycle Area : The area of the first period of waveform on the screen and the unit is
voltage-second. The area above the zero reference (namely the vertical offset) is positive and
the area below the zero reference is negative. The area measured is the algebraic sum of the
area of the whole period waveform.
Note: When the waveform on the screen is less than a period, the period area measured is 0.
How to Measure with Cursors
Push the Cursor button to turn cursors on and display the cursor menu. Push it again to turn
cursors off.
The Cursor Measurement for normal mode:
The description of the cursor menu is shown as the following table:
Function
Menu
Setting
Description
Type
Voltage
Time
Time&Voltage
AutoCursr
Display the voltage measurement cursor and menu.
Display the time measurement cursor and menu.
Display the time and voltage measurement cursor and
menu.
The horizontal cursors are set as the intersections of the
vertical cursors and the waveform
Line Type
(Time&Volt
age type)
Time
Voltage
Makes the vertical cursors active.
Makes the horizontal cursors active.