User Manual

View→ Page Transform: Switch the page orientation between landscape and portrait 
  Whole page: Display the full page on the screen 
  Face size: Display in actual size 
  Default Scale...: Display in a specified zoom factor 
Show  Wave  Background:  Check  to  display  wave  background  color;  uncheck  to  display 
preview paper background 
Set  Preview  Page  Background: Click to display color picker  dialog box to set the color of 
preview page background 
Save Image 
Click to save screen shot as an image file in png, bmp or gif format. 
Save/Refer 
This  function  allows  to  store  8  reference  waveforms.  These  waveforms  can  be  displayed  with  the 
current waveform simultaneously. The recalled waveform can´t be adjusted. The source can be CH1, 
CH2 or Math. 
To save the waveform of CH1 into object "a" and recall it, the steps are as follows: 
1.  Choose CH1 as source. 
2.  Choose "To Object" as "a". 
3.  Fillin "Object Rename", such as "sine". Click "OK". The object name is renamed as "a (sine)". This 
step  can be skipped. 
4.  Click "save". 
5.  Select "a (sine)" from the list, tick "Show", the reference waveform will appear on the screen. The 
object  name  and relevant  information  will be displayed  on the top left. You can  continue  to choose 
another  reference  waveform  to  display.  Click  "Remove  all"  to  clear  all  reference  waveforms  on  the 
screen. 
If the chosen object has no stored waveform, it will display "No Saved" below. 
Pause&Export 
Export the waveform into a file in specified format according to current record length. You can choose 
bin, txt, csv, xls as the  format.  You can  also  click   icon on  the  user  interface  to save it  in the 
format specified in menu. 
Self Cal 
The self-calibration can make the oscilloscope reach the optimum condition rapidly to obtain the most 
accurate measurement value. You can carry out this application program at any time. This program 
must be executed whenever the change of ambient temperature is 5°C or over. 
Before  performing  a  self-calibration,  disconnect  all  probes  or  wires  from  the  input  connector.  After 
everything is ready, click "Self Cal". 
Default 
Click "Default" to call out the factory settings. You can also click icon on the user interface. 
Help 
Open this help document. You can press F1 as shortcut key. 
-94- 