User Manual

Hanning window 
Notes for using FFT 
  Select "Scale" to magnify the FFT wave form if necessary. 
  Use  the  default  dB  scale  for  details  of  multiple  frequencies,  even  if  they  have  very  different 
amplitudes. Use the Vrms scale to compare frequencies. 
  DC component or offset can cause incorrect magnitude values of FFT waveform. To minimize the 
DC component, choose AC Coupling on the source signal. 
  To  reduce  random  noise  and  aliased  components  in  repetitive  or  single-shot  events,  set  the 
oscilloscope acquisition mode to average. 
What is Nyquist frequency? 
The Nyquist frequency is the highest frequency that any real-time digitizing oscilloscope can acquire 
without aliasing. This frequency is half of the sample rate. 
Frequencies above the Nyquist frequency will be under sampled, which causes aliasing. So pay more 
attention to the relation between the frequency being sampled and measured. 
Note:  
In FFT mode, only support to set coupling, voltage division and zero position in Channel window, other 
operations are prohibited. If you want to enter other menus, uncheck "FFT" first. 
16. How to zoom the waveform 
Click 
to show Function menu, choose “Horizon”. 
16.1. Main Time Base 
The setting of the horizontal main time base is used to display the waveform. 
16.2. Assist Set 
A window area is defined by two cursors, which  will be expanded to the full screen size  in Window 
Expansion. 
Choose W value from the combo box to adjust the size of this window area. Click   to show the 
slider bar. You can adjust the horizontal position of the area in main window by it. 
Note: In Assist Set status, the time base and horizontal trigger position cannot be set. 
-89- 