User Manual

Pulse Width Trigger 
Choose Mode as "Pulse". Pulse trigger occurs according to the width of pulse. The abnormal signals 
can be detected through setting up the pulse width condition. 
1.  Select pulse width condition and set time. 
2.  Set trigger level. 
3.   Set Trig Mode and Hold Off. About the setting method, refer to Edge Trigger. 
Alternate trigger 
Trigger signal comes from two vertical channels when alternate trigger is on. This mode is used to 
observe two unrelated signals. You can choose different trigger modes for different channels. The 
options are as follows: edge, video, pulse or slope. 
Set in Trigger menu: 
1.   In Trigger menu, choose "Alternate" (the chosen label has   a mark). 
2.  Choose Source. 
3.  Choose Mode. 
About the setting method of Mode and parameters, refer to Single trigger. 
Note: In alternate trigger, only one channel at most can be set as Vide mode. You cannot choose 
Video in both channels. 
Set in Trigger window: 
Instruction of Trigger mode icon in Trigger window: 
 Rise in Edge mod 
 Fall in Edge mode 
 Synchronic trigger in video lin 
 Synchronic trigger in video field 
 Synchronic trigger in video odd file 
 Synchronic trigger in video even field 
 Rising in Slop 
 Falling in Slope 
 +Pulse Widt 
 -Pulse Width 
-77- 