User Manual

4.1. How to Implement the Function Inspection  
Make  a  fast  function  check  to  verify  the  normal  operation  of  the  instrument.  First,  connect  the 
oscilloscope with PC successfully, and then do the following steps:  
1.  Click   in the software interface to restore to factory settings. The default attenuation coefficient 
set value of the probe in the software is 10X. 
2.  Set  the  Switch  in  the  Oscilloscope  Probe  as  10X  and  Connect  the  Oscilloscope  with  CH1 
Channel. 
Align the slot in the probe with the plug in the CH1 connector BNC, and then tighten the probe 
with rotating it to the right side. 
Connect the probe tip and the ground clamp to the connector of the probe compensator. 
3.  Click   on the right top of the software interface to auto set. 
The square wave of 1 KHz frequency and 3.3 peak-peak value will be displayed. 
4.  Check CH2 by repeating Step 2 and Step 3. 
4.2. How to Implement the Probe Compensation  
When connect the probe with any input channel for the first time, make this adjustment to match the 
probe  with  the  input  channel.  The  probe  which  is  not  compensated  or  presents  a  compensation 
deviation will result in the measuring error or mistake. For adjusting the probe compensation, please 
carry out the following steps: 
1.  Click    in the software interface to show the Home menu. Choose "Channel", set the "Probe 
Rate"  in the menu as 10. Set the  switch  in  the  probe  as 10X  and  connect  the  probe  with the CH1 
channel. If a probe hook tip is used, ensure that it keeps in close touch  with the probe. Connect the 
probe tip with the signal connector of the probe compensator and connect the reference wire clamp 
with  the  ground  wire  connector  of  the  probe  connector,  and  then  click    on  the  right  top  of  the 
software interface to auto set.  
The set values of the probe switch are 1X and 10X 
Figure 0-2 Attenuation Switch 
-58- 