User Manual

Ground Clip
Signal Input
Electrical Outlet
Probe
Oscilloscope AC Adapter
The diagram of the ground wire connection when the oscilloscope powered by PC USB is connected 
to the AC-powered PC through the USB ports: 
Ground Clip
Signal Input
Oscilloscope
(PC USB power)
PC Electrical OutletProbe
USB Cable
It is not allowed to measure AC power when the oscilloscope is AC powered through the adapter, or 
when the oscilloscope powered by PC USB is connected to the AC-powered PC through the ports. 
  Warning: 
To avoid fire or electrical shock if a test tool input is connected to more 42V peak (30Vrms) or 
on circuits of more than 4800VA: 
  Use only insulated probes and adapter supplied with the test tool, or indicated as suitable for the 
Oscilloscope. 
  Before use, inspect probes and accessories for mechanical damage and replace when damaged.  
  Remove all probes and accessories that are not in use.  
  Do not apply voltages that differ more than 400 V from earth ground to any input when measuring 
in a CAT II environment. 
  Do  not  apply  voltages  that  differ  more  than  400  V  from  each  other  to  the  isolated  inputs  when 
measuring in a CAT II environment. 
  Do  not apply  input  voltages  above  the rating  of  the  instrument  Use  caution  when  using  1:1  test 
leads because the probe tip voltage will be directly transmitted to the Oscilloscope. 
  Do not touch the exposed part of metal BNC. 
  Do not insert metal objects into connectors. 
  Always use the Oscilloscope only in the manner specified. 
  Voltage  ratings  that  are  mentioned  in  the  warning  are  giver  as  limits  for  "working  voltage".  The 
represent  V  ac  rms  (50-60Hz)  for  AC  sine  wave  applications  and  as  V  dc  for  DC  applications. 
Overvoltage  Category  II  refers  to  local  level,  which  is  applicable  for  appliances  and  portable 
equipment. 
-55- 